Use this URL to cite or link to this record in EThOS: http://ethos.bl.uk/OrderDetails.do?uin=uk.bl.ethos.594965
Title: The use of pseudo-random binary sequences in gas chromatography
Author: Moss, Geoffrey Charles
Awarding Body: University of Warwick
Current Institution: University of Warwick
Date of Award: 1971
Availability of Full Text:
Access from EThOS:
Access from Institution:
Abstract:
Techniques of conventional chromatography do not always give convenient results. Analyses are often contaminated with detector noise and baseline drift. Continuous analysis is not possible and anomalous detector response can occur. This report proposes a method which overcomes these difficulties to a certain degree. A pseudo- random sequence controls the injections into a conventional column. Cross-correlation between input and output yields the chromatogram to within a constant factor. Sample valve and detector must be operated in a linear fashion. The resultant data analysis requires digital computation, but this can be carried out on-line.
Supervisor: Not available Sponsor: Not available
Qualification Name: Thesis (Ph.D.) Qualification Level: Doctoral
EThOS ID: uk.bl.ethos.594965  DOI: Not available
Keywords: TK Electrical engineering. Electronics Nuclear engineering