Use this URL to cite or link to this record in EThOS: http://ethos.bl.uk/OrderDetails.do?uin=uk.bl.ethos.559531
Title: Radio frequency probes for ultra-high field magnetic resonance imaging
Author: Lee, Daniel
Awarding Body: University of Nottingham
Current Institution: University of Nottingham
Date of Award: 2011
Availability of Full Text:
Access from EThOS:
Access from Institution:
Abstract:
This thesis describes the design, construction and testing of a dome coil. The dome coil is hemispherical in shape and is intended to be used within a set of hemispherical gradient coils in a seven tesla magnetic resonance imaging magnet. The dome coil has eight independent elements and is designed to be used for parallel transmission and reception. It is shown that the dome coil produces less specific absorption rate than a conventional birdcage coil and is suitable for head imaging. A study of travelling wave magnetic resonance imaging is also presented. In this study two different methods of generating a travelling wave (using a patch antenna and an end-fire helix antenna) are compared, and the power requirements and specific absorption rate of the travelling wave approach are considered. It is concluded that travelling wave magnetic resonance imaging is best performed using a local receive coil and with the travelling wave antenna placed at least 50 cm from the subject. All theory relevant to the design, construction, testing and use of these coils and antennas is also presented.
Supervisor: Not available Sponsor: Not available
Qualification Name: Thesis (Ph.D.) Qualification Level: Doctoral
EThOS ID: uk.bl.ethos.559531  DOI: Not available
Keywords: QC501 Electricity and magnetism
Share: